Fig. 5 | Beni-Suef University Journal of Basic and Applied Sciences

Fig. 5

From: Aerosol delivery through high flow nasal cannula compared to biphasic positive airway pressure, at two different pressure: an in-vitro study

Fig. 5

Modified from ElHansy et al. [24]. SOLO vibrating mesh nebulizer was inserted in the ventilation circuit in the inspiratory limb of the BiPAP (before the Y piece of the dual limb ventilation circuit). Cooled Anderson Cascade Impactor (inhalation flow rate of 15 L min−1) was connected to the Y piece. The bi-level ventilator was adjusted at two different pressures; IPAP/EPAP 20/5 cm water and IPAP/EPAP 10/5 cm water

Schematic diagram showing the measurement of the particle size distribution, by Anderson Cascade Impactor, using BiPAP mode.