Skip to main content

Table 2 Analytical parameters for the quantification of TIG in intravenous infusion solution

From: In situ green analytical methods for the rapid and sensitive determination of a newly launched orphan anticancer drug; Tigecycline in infusion bags: comparative study

Parameters

CZE method

Spectrophotometric methods

Method I

Method II

Method III

ΔA

ΔD1

Linearity range (µg/mL)

5–300

20–120

40–140

40–140

20–140

QL (µg/mL)

3.25

11.46

17.02

15.58

24.41

DL (µg/mL)

0.98

3.78

5.62

5.14

8.06

Intercept (a)

6.34 × 10−2

1.33 × 10−2

4.38 × 10−2

5.56 × 10−2

2.78 × 10−1

Slope (b)

5.01 × 10−1

3.43 × 10−3

9.13 × 10−3

4.71 × 10−3

2.91 × 10−3

Correlation coefficient (r)

0.9999

0.9997

0.9995

0.9994

0.9990

Sa

3.00 × 10−1

3.84 × 10−3

2.06 × 10−2

9.73 × 10−3

6.91 × 10−3

Sb

2.18 × 10−3

4.74 × 10−5

2.02 × 10−4

9.55 × 10−5

7.37 × 10−5

Sb2

4.74 × 10−6

2.25 × 10−9

4.08 × 10−8

9.12 × 10−9

5.43 × 10−9

Sy/x

6.25 × 10−1

4.00 × 10−3

1.55 × 10−2

7.34 × 10−3

7.10 × 10−3

F

52,936.46

5237.07

2042.96

2437.61

1556.59

Significance F

7.73 × 10−15

5.81 × 10−6

2.38 × 10−5

1.83 × 10−5

3.58 × 10−5