Table 1 Comparison between supplied doses of salbutamol and the inhalation filter in micrograms at different oxygen flow in L/min using mouthpiece and facemask

From: Effect of oxygen flow on aerosol delivery from a vibrating mesh nebulizer with a holding chamber

O2 flow rate (L/min)

Mouthpiece

Facemask

Delivered dose

Nebulizer residual

Amount deposited at holding chamber

Delivered dose

Nebulizer residual

Amount deposited at holding chamber

0

2153.9 ± 495.1

309.5 ± 87.3

443.2 ± 244.9

2136.3 ± 285.5

201.3 ± 165.4

292.2 ± 245.6

1

2184.6 ± 393.4

507.9 ± 288.4

108.0 ± 99.0

2164.7 ± 477.4

271.7 ± 171.2

142.1 ± 116.3

2

2445.7 ± 315.4

305.3 ± 245.4

45.2 ± 29.6

2230.6 ± 105.0

365.8 ± 189.6

170.3 ± 90.5

3

2190.1 ± 296.5

555.6 ± 282.3

197.1 ± 135.5

2836.2 ± 188.6

228.9 ± 119.7

149.3 ± 161.0

4

2158.6 ± 201.8

400.1 ± 314.1

150.9 ± 137.1

2570.2 ± 76.1

470.6 ± 140.6

149.4 ± 57.1

5

2076.2 ± 506.8

408.7 ± 226.7

139.5 ± 86.2

2284.2 ± 246.5

525.1 ± 297.0

115.9 ± 37.8

6

1814.5 ± 162.3

449.2 ± 198.8

35.1 ± 25.6

2128.1 ± 243.1

567.5 ± 294.8

154.0 ± 41.7